Autoclave Sterilization Temperature and Time Pressure Chart?

STERILIZER    TEMPERATURE    PRESSURE    TIME
Steam autoclave    121 C (250 F)    15psi    15min
unwrapped items    132 C (270 F)    30psi    3min
lightly wrapped items    132 C (270 F)    30psi    8min
heavily wrapped items    132 C (270 F)    30psi    10min
Dry heat wrapped    170 C (340 F)        60 min
     160 C (340 F)         120min
150 C (300F)        150min
140 C (285F)        180min
121 C (250F)         12hrs
Dry heat (rapid flow)    190 C (375F)        6min
unwrapped items   
Dry heat (rapid flow)    190 C  (375 F)        12min
packaged items   
Chemical vapor    132 C (270 F)    20-40 psi    20min
Ethylene oxide    Ambient        8-10 hours