How Do I maintain my Ultrasonic Cleaner Sydney System?

1. Location Locate the cleaner in an area where you have plenty of room to stage your parts prior to cleaning and where you have room to stack the clean parts after cleaning. Do not block the generator airflow as this will overheat the generator and cause a loss of cleaning action and damage the electronics. If your generator is separate from the tank make sure that it wont get wet when you are cleaning parts.

2. Foil Test Get a base line for performance for your new tank; Even if you cant afford a probe to test and record and ultrasonics tank performance you can do a simple aluminum foil test to determine your tanks initial performance and keep this record in a book to compare to future records. To perform this test you will need some aluminum foil Note: this test can contaminate a critical cleaning system DO NOT USE!